Rule 44:03 RADIATION SAFETY
ARTICLE 44:03
RADIATION SAFETY
Chapter
44:03:01 Radiation control.
Rule 44:03:01 RADIATION CONTROL
Rule 44:03:01:01 Definitions.
44:03:01:01. Definitions.
Terms defined by SDCL 34-21-2 have the same meaning when used in this chapter. In addition, the terms used in this chapter mean:
(1) "Added
filtration," any filtration that is in addition to the inherent
filtration;
(2) "Aluminum
equivalent," the thickness of type 1100 aluminum alloy affording the same
attenuation, under specified conditions, as the material in question;
(3) "Assembler,"
any person engaged in the business of assembling, replacing, or installing one
or more components into an X ray system or subsystem. The term includes the
owner of an X ray system or the owner's employee or agent who assembles
components into an X ray system that is subsequently used to provide
professional or commercial services;
(4) "Automatic
exposure control (AEC)," a device that automatically controls one or more
technique factors in order to obtain at a preselected location a required
quantity of radiation. The term includes devices such as phototimes and ion
chambers;
(5) "C-arm X ray
system," and X ray system where the image receptor and X ray tube housing
assembly are connected by a common mechanical support system in order to
maintain a desired spatial relationship and which is designed to allow a change
in the projection of the beam through the patient without a change in the
position of the patient;
(6) "Certified
components," components of X ray systems that are subject to regulations
by the Food and Drug Administration under the Radiation Control for Health and
Safety Act of 1968, Pub. L. No. 90-602;
(7) "Certified
system," any X ray system that has one or more certified components;
(8) Computed
tomography" or "CT," the production of a tomogram by the
acquisition and computer processing of X ray transmission data;
(9) "Control
panel," that part of the X ray control upon which are mounted the
switches, knobs, push buttons, and other hardware necessary for manually
setting the technique factors;
(10) "Cooling
curve," the graphical relationship between heat units stored and cooling
time;
(11) "CRT,"
cathode ray tube in which cathode rays are used to produce an image on a
fluorescent screen;
(12) "Dead-man
switch," a switch so constructed that a circuit closing contact can be
maintained only by continuous pressure on the switch by the operator;
(13) "Department,"
the Department of Health;
(14) "Diagnostic X ray
system," an X ray system designed for irradiation of any part of the human
or animal body for the purpose of diagnosis or visualization;
(15) "Diagnostic X ray
imaging system," an assemblage of components for the generation, emission,
and reception of X ray and the transformation, storage, and visual display of
the resultant X ray image;
(16) "Dose," a
quantity of radiation exposure to the whole anatomy or any portion of the human
or animal body;
(17) "Exposure
survey," an evaluation of the radiation hazards incident to the
production, use, release, disposal, or presence of radiation. When appropriate,
such an evaluation includes a physical survey of materials and equipment and
measurements of levels of radiation or concentration of radioactive material
present;
(18) "Fluoroscopy
imaging assembly," a subsystem in which X ray photons produce a visible
image. The term includes the image receptor such as the image intensifier,
spot-film device, electrical interlocks, if any, and structural material
providing linkage between the image receptor and diagnostic source assembly;
(19) "Gonad
shield," a protective barrier for the testes or ovaries;
(20) Half-value
layer," the thickness of specified material that attenuates the beam of
radiation to an extent such that the exposure rate is reduced by one-half. For
the purpose of this definition, the contribution of all scattered radiation,
other than any which might be present initially in the beam concerned, is
deemed to be excluded;
(21) "Health
arts," those professional disciplines authorized by the laws of South
Dakota (SDCL chapter 36-2) to use X ray or radioactive material in the diagnosis or treatment of human or animal disease;
(22) "Heat unit,"
a unit of energy equal to the product of the peak kilovoltage, milliamps, and
seconds, i.e. kVp x mA x second;
(23) "Image
intensifier," a device, installed in its housing, that instantaneously
converts an X ray pattern into a corresponding light image of higher intensity;
(24) "Image
receptor," any device, such as a fluorescent screen or radiographic film,
that transforms incident X ray photons either into a visible image or into
another form that can be made into a visible image by further transformations;
(25) "Inherent
filtration," the filtration of the useful beam provided by the permanently
installed components of the tube housing assembly;
(26) "kVp," the
maximum value in kilovolts of the potential difference of an X ray generator;
(27) "Lead
equivalent," the thickness of lead affording the same attenuation as the
material in question;
(28) "Leakage
radiation," any radiation coming from within the source housing except for
the useful beam and radiation produced when the exposure switch or timer is not
activated;
(29) "Licensed
practitioner of the healing arts," health professionals for diagnostic or
healing treatment of human and animal maladies licensed by the state of South
Dakota (SDCL chapter 36-2) for the lawful practice of medicine;
(30) "Light
field," that area of the intersection of the light beam from the
beam-limiting device and one of the set of planes parallel to and including the
plane of the image receptor, whose perimeter is the locus of points at which
the illumination is one-fourth of the maximum in the intersection;
(31) "mA,"
milliampere;
(32) "mAs,"
milliampere second;
(33) "Patient,"
an individual or animal subjected to healing arts examination, diagnosis, or
treatment;
(34) "Peak tube
potential," the maximum value of the potential difference across the X ray
tube during an exposure;
(35) "Personnel
monitoring," the use of film badges, pocket chambers, or other devices
worn or carried on individuals for the monitoring of personnel exposures to
radiation;
(36) "Positive beam
limitation," the automatic or semi-automatic adjustment of an X ray beam
to the size of the selected image receptor, whereby exposures cannot be made
without such adjustment;
(37) "Protective
apron," an apron made of radiation absorbing materials used to reduce
radiation exposure;
(38) "Protective
glove," a glove made of radiation absorbing material used to reduce
radiation exposure;
(39) "Qualified expert,"
an individual who possess the knowledge, training, and experience to measure
ionizing radiation, to evaluate safety techniques, and to advise regarding
radiation protection needs including health physicists;
(40) "Qualified
instructor," an individual who possess the knowledge, training, and
experience in the field of radiation to teach fundamentals of radiation safety,
equipment operation, film processing, emergency procedures, personnel
dosimetry, anatomy and physiology, and radiographic positioning;
(41) "Rad," the
special unit of absorbed dose. One rad is equal to an absorbed dose of 100 erg
per gram or 0.01 joule per kilogram;
(42) "Radiation
hazard," a condition under which a person might receive radiation in
excess of the maximum permissible dose;
(43) "Rem," the
special unit of any of the quantities expressed as dose equivalent. The dose
equivalent in rem is equal to the absorbed dose in rad;
(44) "Scattered
radiation," radiation that, during passage through matter, has been
deviated in direction;
(45) "Services,"
may include calibration of radiation-producing machines or instruments,
radiation protection surveys, shielding design, radiological health
consultations, and personnel dosimetry;
(46) "Shielding,"
a protective barrier used to reduce radiation exposure to the required degree.
For the purpose of this term, a primary protective barrier is the material,
excluding filters, placed in the useful beam and a secondary protective barrier
is the material that attenuates stray radiation;
(47) "Source-image
receptor distance" or "SID," the distance from the source to the
center of the input surface of the image receptor;
(48) "Spot film,"
a radiograph that is made during a fluoroscopic examination to permanently
record conditions that exist during that fluoroscopic procedure;
(49) "Spot-film
device," a device intended to transport or position, or both, a
radiographic image receptor between the X ray source and fluoroscopic image
receptor. The term includes a device intended to hold a cassette over the input
end of an image intensifier for the purpose of making a radiograph;
(50) "Stray
radiation," the sum of leakage and scattered radiation;
(51) "Target,"
the point at which an X ray is produced;
(52) "Technique factors,"
the following conditions of operation:
(a) For capacitor
energy storage equipment, peak tube potential in kV and quantity of charge in
mAs;
(b) For field emission
equipment rated for pulsed operation, peak tube potential in kV, and number of X
ray pulses;
(c) For CT X ray
systems designed for pulsed operation, peak tube potential in kV, scan time in
seconds, and either tube current in mA, X ray pulse width in seconds, and the
number of X ray pulses per scan, or the product of tube current, X ray
pulse width, and the number of X ray pulses in mAs;
(d) For CT X ray
systems not designed for pulsed operation, peak tube potential in kV, and
either tube current in mA and scan time in seconds, or the product of tube
current and exposure time in mAs and the scan time if the scan time and
exposure time are equivalent; and
(e) For all other
equipment, peak tube potential in kV, and either tube current in mA and
exposure time in seconds, or the product of tube current and exposure time in
mAs;
(53) "Tube," an X
ray tube;
(54) "Variable-aperture
beam-limiting device," a beam-limiting device that has capacity for
stepless adjustment of the X ray field size at a given SID;
(55) "X ray exposure
control," a device, switch, button, or other similar by which an operator
initiates or terminates, or both, the radiation exposure. The term may include
such associated equipment as timers and back-up timers;
(56) "X ray
equipment," an X ray system, subsystem, or component of the system;
(57) "X ray
field," that area of the intersection of the useful beam and any one of
the set of planes parallel to and including the plane of the image receptor,
whose perimeter is the locus of points at which the exposure rate is one-fourth
of the maximum in the intersection;
(58) "X ray
system," an assemblage of components for the controlled production of X
rays. The term includes minimally an X ray high-voltage generator, an X ray
control, a tube housing assembly, a beam-limiting device, and the necessary supporting
structures. Any additional component, which functions with the system, is
considered an integral part of the system; and
(59) "X ray
tube," any electron tube which is designed for the conversion of
electrical energy into X ray energy.
Source:
SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980; 26 SDR 96,
effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-11, 34-21-18.
Rule 44:03:01:02 Licensing of radiation-producing devices and materials.
44:03:01:02. Licensing of radiation-producing devices and
materials. Licensing of radiation sources or materials is required for the
production, transport, transfer, receipt, acquisition, possession, use,
storage, or disposal of radiation sources or materials used in the healing
arts. Licensing shall be accomplished using procedures and forms required by
the department.
Source:
SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980; 26 SDR 96,
effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-11, 34-21-12, 34-21-18, 34-21-20.
Rule 44:03:01:03 Transferred.
44:03:01:03. Transferred to § 44:03:01:06.01.
Rule 44:03:01:04 Application for license.
44:03:01:04. Application for license. Each person having
a radiation source, a radiation device, or a radioactive material facility
shall apply for licensure with the department within 30 days after the
installation of the equipment or purchase of material. The license shall be
obtained before the equipment is operated. The application for license shall be
completed on forms furnished by the department and shall contain all the
information required by the form and accompanying instructions. A copy of the current
United States Nuclear Regulatory Commission license must accompany the
application if applying for a radioactive material license.
Source:
SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980; 26 SDR 96,
effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-11, 34-21-18.
Rule 44:03:01:05 Repealed.
44:03:01:05. Renewal of registration.Repealed.
Source:
SL 1975, ch 16, § 1; repealed, 6 SDR 93, effective July 1, 1980.
Rule 44:03:01:06 Annual license renewal.
44:03:01:06. Annual license renewal. The licensee shall
notify the department in writing within 30 days after any change which makes
the location or other information on machines, devices, or other radiation
sources no longer accurate. The license shall be renewed annually during the
month of January on forms supplied by the department.
Source:
SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980; 26 SDR 96,
effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-22.
Rule 44:03:01:06.01 Exemption from radiation licensing.
44:03:01:06.01. Exemption from radiation licensing.
Electronic equipment that produces radiation incidental to its operation for
other purposes is exempt from licensing if the radiation dose equivalent rate
averaged over an area of ten square centimeters does not exceed 0.5 millirem per hour at five centimeters from any accessible
surface of the equipment. Domestic television receivers and CRTs are
exempt from the licensing requirement.
Source:
SL 1975, ch 16, § 1; transferred from § 44:03:01:03, 6 SDR 93,
effective July 1, 1980; 26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-20.
Rule 44:03:01:06.02 Repealed.
44:03:01:06.02 Annual license fee. Repealed.
Source:
31 SDR 62, effective November 7, 2004; repealed, 35 SDR 305, effective July 1,
2009.
Rule 44:03:01:07 Repealed.
44:03:01:07. Out-of-state registrable items.Repealed.
Source:
SL 1975, ch 16, § 1; repealed, 6 SDR 93, effective July 1, 1980.
Rule 44:03:01:07.01 Licensing of radiation producing devices for temporary use.
44:03:01:07.01. Licensing of radiation producing devices for
temporary use. Before any radiation machine is brought into the state for
any temporary use, the person proposing to bring such machine into the state
shall give written notice to the department at least ten working days before
such machine is to be used in the state. A license is required and the person
shall comply with the provisions of this chapter. The notice shall include:
(1) The type of radiation
machine;
(2) The nature, duration, and
scope of use; and
(3) The exact location or
locations where the radiation machine is to be used.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-20.
Rule 44:03:01:08 Prohibited uses of radiation.
44:03:01:08. Prohibited uses of radiation. No person may
be exposed to diagnostic or therapeutic radiation except for healing arts
purposes and only if the exposure has been authorized by a licensed
practitioner of the healing arts. No person may be exposed to the useful beam
for non-healing arts training, instruction, or demonstration, or other
purposes. The following radiation producing equipment may not be used and the
following specified procedures may not be performed:
(1) Fluoroscopic devices
for fitting shoes;
(2) Photofluorographic
equipment;
(3) Dental fluoroscopic
imaging assemblies;
(4) Hand-held radiographic
or fluoroscopic imaging devices, except for intra-oral radiographic imaging
devices;
(5) The use of fluoroscopy
for positioning a patient for general radiographic imaging, except for
radiation therapy simulators;
(6) The use of fluoroscopy
and c-arm fluoroscopes by a person other than a licensed practitioner of the
healing arts unless under the supervision of a licensed practitioner of the
healing arts;
(7) The use of direct
exposure X ray film (without intensifying screens) for routine diagnostic
procedures other than intraoral dental radiography, therapeutic portal imaging,
and industrial radiography;
(8) Nonimage intensified
fluoroscopic X ray equipment; or
(9) The use of X ray
equipment for mammography unless specifically designed by the manufacturer for
the imaging of the breast.
Source:
SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980; 26 SDR 96,
effective January 23, 2000; 35 SDR 47, effective September 8, 2008.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-23.
Rule 44:03:01:08.01 Repealed.
44:03:01:08.01. Equipment standards.Repealed.
Source:
SL 1975, ch 16, § 1; transferred from § 44:03:01:13, 6 SDR 93,
effective July 1, 1980; repealed, 26 SDR 96, effective January 23, 2000.
Rule 44:03:01:08.02 Equipment standards for medical diagnostic X ray machines.
44:03:01:08.02. Equipment standards for medical diagnostic X ray
machines. The standards for any medical diagnostic X
ray machine are as follows:
(1) The control panel
containing the main power switch shall bear the warning statement, legible and
accessible to view: "WARNING: This X ray unit may be dangerous to patient
and operator unless safe exposure factors and operating instructions are
observed.";
(2) If the machine contains
a battery-powered X ray generator, a visual means
shall be provided on the control panel to indicate if the battery is in a state
of charge adequate for proper operation;
(3) Any leakage radiation
from the diagnostic source assembly measured at a distance of one meter in any
direction from the source may not exceed 100 milliroentgens in one hour;
(4) The filtration or beam
quality is considered adequate if the total filtration in the beam is not less
than the following table:
|
Operating
Voltage vs. Total Filtration Required
|
(Total filtration = inherent plus added)
|
|
Operating
Voltage
|
|
|
(Peak
kilovolt) (kVp)
|
(Millimeters
aluminum equivalent)
|
|
|
|
|
Below
50
|
0.5
millimeters
|
|
50
-- 70
|
1.5
millimeters
|
|
Above
70
|
2.5
millimeters
|
(5) A variable, positive
beam limitation with rectangular area and a light-defining device shall be
provided for all fixed X ray machines. The X ray beam
dimensions may not exceed the size of film used by greater than two percent of
SID on any side. The machine shall include a means to align the center of the X
ray field with respect to the center of the image receptor to within two
percent of the SID. If a light localizer is used to define the X ray field, it shall provide an average illumination of not
less than 160 lux (15.0 foot-candles) at 100 centimeters or at the maximum SID,
whichever is less. The average illumination shall be based upon measurements
made in the approximate center of each quadrant of the light field. Exempt is
any X ray machine that is designed with all parameters
fixed, including alignment, source-to-distance, and technique factors;
(6) The machine shall
include a device to terminate the exposure after a preset time. The accuracy of
such a device shall be within five percent of the time set for machines
manufactured on or after August 1, 1974, and within ten percent of the time set
for machines manufactured before August 1, 1974;
(7) Any deviation of a
measured technique factor from an indicated value of kVp may not exceed any
limit specified for that system by its manufacturer or, in the absence of any
manufacturer's specifications, the deviation may not
exceed ten percent of the indicated value for kVp;
(8) The coefficient of
variation may not exceed 0.10 when all technique factors are held constant.
This requirement is met if, when four exposures are made at identical technique
factors, the value of the average exposure (E) is greater than or equal to five
times the maximum exposure (E max) minus the minimum exposure (E min), i.e., E
> 5(E max - E min);
(9) MA/mAs linearity
requirements apply if the equipment is being operated on a power supply as
specified by the manufacturer for any fixed X ray tube
potential within the range of 40 percent to 100 percent of the maximum rated:
(a) For equipment
having independent selection of X ray tube current
(mA), the average ratios (X1) exposure to the indicated
milliampere-seconds product, in units of coulombs per kilograms per milliampere
second (or milliroentgen per milliampere-seconds), obtained at any two
consecutive tube current settings may not differ by more than ten hundredths
times their sum:
X1 - X2 <
0.10 (X1 + Z2)
where
X1 and X2 are the average values obtained by two mAs
selector settings, or at two settings differing by no more than a factor of two
where the mAs selector provides continuous selection;
(b) For
equipment having a combined X ray tube current-exposure time product (mAs)
selector, but not a separate tube current (mA) selector, the average ratios (X1)
of exposure to the indicated milliampere-seconds product, in units of coulombs
per kilogram per milliampere second (or mR/mAs), obtained at any two
consecutive mAs selector settings may not differ by more than ten hundredths
times their sum:
X1 - X2 <
0.10 (X1 + Z2)
where
X1 and X2 are the average values obtained by two mAs
selector settings, or at two settings differing by no more than a factor of two
where the mAs selector provides continuous selection;
(10) If two or more
radiographic tubes are controlled by one exposure switch, the tube that has
been selected shall be clearly indicated prior to initiation of the exposure.
This indication shall be on the X ray control panel and also at or near the
tube housing assembly which has been selected;
(11) The tube housing
assembly supports shall be adjusted so that the tube housing assembly remains
stable during an exposure unless tube housing movement is a designed function
of the X ray system;
(12) Any diagnostic X ray
system and its associated components used on humans and certified pursuant to
the Federal X Ray Equipment Performance Standard (21 C.F.R. Part 1020) as of
January 1, 1998, shall be maintained in compliance with applicable requirements
of that standard; and
(13) All
position locking, holding, and centering devices on the machine shall function
as intended by the manufacturer.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:08.03 Equipment standards for medical fluoroscopic X ray machines.
44:03:01:08.03. Equipment standards for medical fluoroscopic X
ray machines. Fluoroscopic X ray equipment shall be image intensified and
the standards are as follows:
(1) The filtration or beam
quality is considered adequate if the total in the beam is not less than the
table below:
|
Operating
Voltage vs. Total Filtration Required
|
(Total filtration = inherent plus added)
|
|
Operating
Voltage
|
|
|
(Peak
kilovolt) (kVp)
|
(Millimeters
aluminum equivalent)
|
|
|
|
|
Below
50
|
0.5
millimeters
|
|
50
-- 70
|
1.5
millimeters
|
|
Above
70
|
2.5
millimeters
|
(2) A manually reset,
cumulative timing device shall be used which must either indicate elapsed time
by an audible signal or turn off the apparatus if the total exposure exceeds a
predetermined limit in one or a series of exposures. The device shall have a
maximum range of five minutes;
(3) Any exposure to the
operator's eyes above the screen and to the operator's waist behind the leaded
drapes may not exceed 50 milliroentgens per hour;
(4) For routine
fluoroscopy, the tabletop exposure may not exceed:
(a) Ten roentgens per
minute with automatic exposure rate control;
(b) Twenty roentgens
per minute with optional high level control provided. A continuous signal
audible to the operator shall indicate that the high level control is being
employed; or
(c) Five roentgens per
minute with fluoroscopic equipment without automatic exposure control;
(5) The fluoroscopic X ray field may not exceed the visible area of the image
receptor by more than four percent of the SID; and
(6) A dead-man switch shall
control the fluoroscopic device.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:08.04 Equipment standards for medical fluoroscopic spot film devices.
44:03:01:08.04. Equipment standards for medical fluoroscopic
spot film devices. Any medical fluoroscopic spot film device shall meet the
following requirements:
(1) The device shall
provide for the adjustment to the size of the spot film selected that is
between the source and the patient. Such an adjustment shall be automatically
accomplished except when the X ray field size is smaller than the selected
portion of the film; and
(2) The center of the X ray field shall be aligned with the center of the selected
portion of the film to within two percent of the SID.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:08.05 Periodic measurement of medical fluoroscopic spot film devices.
44:03:01:08.05. Periodic measurement of medical fluoroscopic
spot film devices. Periodic measurement of the entrance radiation exposure
rate shall be performed by a qualified expert for both typical and maximum
values. Such measurements shall be made triennially or after any maintenance of
the system which might affect the radiation exposure rate. Results of these
measurements shall be posted where any operator may have ready access to such
results while using the fluoroscope. Results of the measurements shall include
the roentgen per minute, as well as the technique factors used to determine
such results. The name of the person performing the measurements and the date
the measurements were performed shall be included in the results.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:08.06 Equipment standards for dental X ray machines.
44:03:01:08.06. Equipment
standards for dental X ray equipment. The standards for dental X ray
equipment are as follows:
(1) The leakage radiation
from the diagnostic source assembly measured at a distance of one meter in any
direction from the source may not exceed 100 milliroentgens in one hour;
(2) The filtration or beam
quality is considered adequate if the total in the beam is not less than the
table below:
|
Operating
Voltage vs. Total Filtration Required
|
(Total filtration = inherent plus added)
|
|
|
|
|
Operating
Voltage
|
(Millimeters
aluminum equivalent)
|
|
|
|
|
Below
70
|
1.5
millimeters
|
|
Above
70
|
2.5
millimeters
|
(3) Collimation of the beam
shall be restricted to a maximum of three inches in diameter and may not have a
pointed cone;
(4) Time accuracy shall be
within five percent of the time set for any X ray equipment installed on or
after August 1, 1974, and ten percent of time set for any equipment installed
before August 1, 1974;
(5) Any deviation of a
measured technique factor from an indicated value for kVp may not exceed any
limit specified for that system by its manufacturer or, in the absence of any
manufacturer's specifications, the deviation may not exceed ten percent of the
indicated value for kVp;
(6) The exposure switch
shall be located so the operator can stand at least six feet from the useful
beam. If sufficient shielding is provided to protect the operator from stray
radiation, the exposure switch may be located closer;
(7) The target to skin
distance shall be at least seven inches;
(8) Any dental X ray
equipment must operate with a kilovoltage of 60 kVp or higher;
(9) Any dental X ray
machine must be maintained within manufacturer's specifications and
recommendations.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:08.07 Equipment standards for mobile X ray equipment.
44:03:01:08.07. Equipment standards for mobile X ray equipment.
Mobile X ray equipment shall meet the standards of
§ 44:03:01:08.02 and as follows:
(1) The exposure switch
shall be located so the operator can stand at least six feet from the useful beam.
If sufficient shielding is provided to protect the operator from stray
radiation, the exposure switch may be located closer; and
(2) Any mobile medical
radiographic equipment shall have a spacer to limit the target-to-skin distance
to at least 12 inches.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:08.08 Equipment standards for computed tomography systems.
44:03:01:08.08. Equipment standards for computed tomography
systems. The standards for computed tomography systems are as follows:
(1) A visible signal must
indicate when the X ray exposure has been terminated.
The operator must be able to terminate the X ray exposure at any time during a
scan or series of scans under CT X ray system control of greater than one-half
second duration;
(2) For any single slice
tomogram system, a means must be provided to permit visual determination of the
tomographic plane or a reference plane offset from the tomographic plane. For
any multiple slice tomogram system, a means must be provided to permit visual
determination of the location of a reference plane. This reference plane can be
offset from the location of the tomographic planes. If a device using a light
source is used to satisfy this subdivision, the light source must provide
illumination levels of not less than 160 lux (15.0 foot-candles) above the room
ambient illumination level;
(3) The X
ray control and gantry must visually indicate whenever X rays are produced and,
if applicable, whether the shutter is open or closed. Any emergency button or
switch must be clearly labeled as to its function. A means shall be provided to
require operator initiation of each individual scan or series of scans;
(4) The CT X ray system
shall be designed to indicate the CT conditions of operation to be used during
a scan or a scan sequence prior to the initiation of a scan or a scan sequence.
On equipment having any of these conditions of operation at fixed values, this
requirement may be met by permanent markings. Indication of CT conditions of
operation shall be visible from any position from which scan initiation is
possible;
(5) The system shall
perform in such a manner that the radiation produced adjacent to the tube
housing assembly, including the tube port, during periods of time scans are not
being performed does not exceed 100 milliroentgen in one hour;
(6) For CT X ray systems
containing a gantry manufactured after September 3, 1985:
(a) The total error in
the indicated location of the tomographic plane or reference plane may not
exceed five millimeters;
(b) If the X ray
production period is less than one-half second, the indication of X ray production shall be actuated for at least one-half
second. Any indicator at or near the gantry shall be discernible from any point
external to the patient opening where insertion of any part of the human body
into the primary beam is possible; and
(c) The deviation of
indicated scan increment versus actual increment may not exceed plus or minus
one millimeter with any mass from 0 to 100 kilograms, inclusive, resting on the
support device. The patient support device shall be incremented from a typical
starting position to the maximum incremented distance or 30 centimeters,
whichever is less, and then returned to the starting position. Measurement of
actual versus indicated scan increments may be taken anywhere along this line
travel;
(7) The system must provide
for two-way oral communication between the patient and the operator at the
control panel;
(8) Leaded windows,
mirrors, closed-circuit television, or an equivalent shall be provided to
permit continuous observation of the patient during irradiation and shall be so
located that the operator can observe the patient from the control panel;
(9) If the primary viewing
system is by electronic means, an alternate viewing system, which may be
electronic, shall be available for use in the event of failure of the primary
viewing system.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:08.09 Hand-held intra-oral radiographic imaging device reporting.
44:03:01:08.09. Hand-held intra-oral radiographic imaging device
reporting. If any hand-held intra-oral radiographic device is damaged or
lost, licensee shall notify the department of the damage or loss within 48
hours. If a device is damaged, licensee shall take the device immediately out
of service and may not return the device to service until the device is
repaired and tested for proper operation. Licensee shall maintain documentation
at the facility that the device has been repaired, tested, and is safe to be
placed back in operation.
Source:
35 SDR 47, effective September 8, 2008.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-23.
Rule 44:03:01:09 Evaluation and correction of hazards.
44:03:01:09. Evaluation and correction of hazards.
Subsequent to the evaluation of radiation hazards, the department may issue
orders for correction of these hazards.
Source:
SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980; 26 SDR 96,
effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-12.
Rule 44:03:01:10 General safety provisions to protect persons from radiation exposures.
44:03:01:10. General safety provisions to protect persons
from radiation exposures. The licensee shall be responsible for directing
the operation of any X ray system under the licensee's
administrative control. The licensee or the licensee's agent shall provide
that:
(1) No X
ray system may be operated for diagnostic purpose unless the system meets the
provisions of this chapter;
(2) Any person who is
operating the X ray system is adequately instructed in
the safe operating procedures and competent in the safe use of the equipment
commensurate with the size, scope, and nature of the service. Any such person shall be instructed and demonstrate competence in subjects
outlined in § 44:03:01:14.01;
(3) Procedures and
auxiliary equipment designed to minimize patient and personnel exposure
commensurate with the needed diagnostic information is
utilized. The speed of film or screen and film combinations shall be the
fastest speed consistent with the diagnostic objective of the examinations. The
radiation exposure to the patient shall be the minimum exposure required to
produce images of good diagnostic quality;
(4) A technique chart or
manual is located in the vicinity of the control panel
of each machine that specifies, for all diagnostic examinations performed with
that system, the following information:
(a) The technique
factors to be used that are specific to a patient's anatomical part, size, or
age (for pediatrics), except for any system that has only automatic exposure
controls;
(b) The type of
film-screen combination to be used;
(c) The type of grid
to be used, if any;
(d) The SID to be used, except for dental and all other fixed SID
radiographic equipment;
(e) The type and
placement of patient shielding to be used;
(f) The routine views
for all procedures done with each machine; and
(g) For mammography,
an indication of kVp/target/filter combination;
(5) A written operating and
safety procedure must be available to each individual who operates radiation
machines. These procedures shall include restrictions for the safe operation of
each radiation machine. The operator shall be able to demonstrate familiarity
with these procedures;
(6) Except for veterinary
facilities, each facility maintains a record containing the patient's name, the
type of examination, the date the examination was performed, and equipment
operator;
(7) Except for patients who
cannot be moved out of the room, only the staff, ancillary personnel or other
persons required for the medical procedure shall be in the room during the
radiographic exposure;
(8) Personnel
monitoring of radiation exposures and records must be maintained by the
licensee of the radiation source. Monitoring with dosimetry devices
shall be required for all persons routinely exposed to radiation in their
occupation. Dental and podiatry offices are exempt from this requirement.
Exposures should not exceed 300 millirems per calendar quarter. Maximum
occupation exposures shall not exceed the limits specified in the following
table:
Rems per Calendar Quarter
|
Whole body; head and trunk; active
blood-forming organs; lenses of eyes; gonads
|
3 3/4
|
|
|
Hands and forearms; feet and ankles
|
18
3/4
|
|
Skin of whole body
|
7 1/2
|
|
|
|
|
|
(9) Radiation sources shall be labeled and caution signs posted to provide a
warning to all persons within the exposure area.
Source: SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980;
26 SDR 96, effective January 23, 2000; 31 SDR 62, effective November 7, 2004.
General Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-25, 34-21-26.
Rule 44:03:01:10.01 Requirements for personal protective devices.
44:03:01:10.01. Requirements for personal protective devices.
Special personal protective devices shall be used to protect eyes, skin, bone,
and certain organs from unnecessary radiation exposure when possible These protective devices must be readily accessible and in
good working condition.
A gonad shield of not less than 0.5
millimeters lead equivalent material must be used for human patients who have
not passed the reproductive age during radiographic procedure in which the
gonads are in the useful beam, unless the shield would interfere with the
diagnostic procedure.
Protective equipment including aprons,
gloves, and shields shall be checked annually for defects, such as holes,
cracks, and tears to assure reliability and integrity. A record of this test
shall be made and maintained. If such defect is found, equipment shall be
replaced or removed from service until repaired;
Mechanical-holding devices shall be
used when the technique permits. The individual holder shall be protected and
no individual may be used routinely to hold film or patients. Written safety
procedures shall indicate the requirements for selecting an individual to be a
holder and the procedure the holder shall follow.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15, 34-21-25.
Rule 44:03:01:10.02 Operator protection requirements.
44:03:01:10.02. Operator protection requirements. Operators
of fixed medical radiographic units shall be within a shielded area large
enough to provide protection from unattenuated scatter or stray radiation
originating from the table or upright cassette holder.
Operators of dental radiographic units
must comply with § 44:03:01:08.06.
A lead glass patient-viewing window,
mirrors, closed circuit television, or an equivalent system must be available
to permit the operator to continuously observe the patient during exposure. If
a patient-viewing window is used, it must be a minimum of one square foot and
must be located at least eighteen inches from the edge of the control booth for
any new construction and any renovation, addition or change in space use of
existing facilities. The exposure switch must be permanently mounted so that it
cannot be conveniently operated outside the shielded area.
For mobile and portable X ray systems to be used less than one week in the same
location, the control must be positioned so that the operator is at least six
feet away from the tube housing and the patient during an exposure and is not
exposed to greater than two millirems in any one hour.
For mobile and portable X ray systems
to be used more than one week in the same location, the operator must be
provided with a movable protective barrier at least 6.5 feet high, 30 inches wide, and a
lead glass viewing window.
Source:
26 SDR 96, effective January 23, 2000; 32 SDR 128, effective January 30, 2006.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:10.03 Reports of incidents involving radiation sources.
44:03:01:10.03. Reports of incidents involving radiation
sources. Any radiation incident considered a potential hazard shall be
reported to the department by the licensee within 24 hours by telephone or fax
at the number shown on the license during normal business hours or the first
workday following a holiday or weekend.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15, 34-21-30.
Rule 44:03:01:10.04 Requirements of X ray film processing and darkroom.
44:03:01:10.04. Requirements of X ray film processing and
darkroom. Each installation using a radiographic X
ray system and using radiographic film shall have available suitable equipment
for handling and processing radiographic film in accordance with the following:
(1) Any manual processing
tank shall be constructed of mechanically rigid corrosion resistant material.
The temperature of solutions in a tank shall be maintained within the range of
60 to 80 degrees Fahrenheit, inclusive. Film shall be developed in accordance
with the time‑temperature relationship recommended by the film
manufacturer. A thermometer and timer shall be utilized to indicate the actual
temperature of the developer and signal the passage of a preset time
appropriate to the developing time required.
(2) Automatic processing of
the film shall be in accordance with the time-temperature relationships
recommended by the film and processor manufacturers. A control program to
maintain the automatic processor operating parameters according to
manufacturer's recommendations must be used;
(3) Film storage shall be
provided and the film stored according to manufacturer's recommendations as to
protection of radiation, heat, humidity, and storage position;
(4) Safelighting in the film
processing-loading area shall be with the filter, bulb wattage, and distances
recommended by the film manufacturer for film emulsion used by the facility.
The safelighting shall prevent a pre-exposed film from increasing in density
greater than 0.10 when exposed for two minutes with the safelights on;
(5) Pass boxes, if
provided, shall be so constructed as to exclude light from the darkroom when
cassettes are placed in or removed from the boxes, and shall incorporate
adequate shielding from stray radiation to prevent exposure of undeveloped
film.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:10.05 Quality assurance program requirements.
44:03:01:10.05. Quality
assurance program requirements. The licensee shall have a written, on-going
quality assurance program specific to the equipment and procedures that are
performed in the facility to ensure consistent high-quality images with minimum
patient exposure. The tests performed for quality control purposes shall be
included in a log containing acceptability limits, results of tests, date,
initials of operator or testing individual, and corrective action taken, if
needed. Tests for film processing shall include temperature, chemical
replacement, processor operating parameters, and darkroom fog, and be performed
on a routine basis. Any quality control test done on diagnostic tubes shall be
done annually and include SID accuracy, X ray and light field alignment, X ray
and bucky alignment, and collimator dial accuracy. All dental intraoral,
panoramic, tomography, and machines that have fixed SID and collimator are
excluded from SID accuracy, X ray and light field alignment, X ray and bucky
alignment, and collimator dial accuracy.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:11 Repealed.
44:03:01:11. Installation of radiation facilities or
equipment.Repealed.
Source:
SL 1975, ch 16, § 1; repealed, 6 SDR 93, effective July 1, 1980.
Rule 44:03:01:11.01 Shielding plan review prior to installation of radiation facilities or equipment.
44:03:01:11.01. Shielding plan review prior to installation of
radiation facilities or equipment. Prior to construction, the floor plans,
shielding specifications, and equipment arrangement of any new installation or
any modification of existing installations utilizing ionizing radiation
machines shall be submitted to the department for review and approval. The
plans shall show at a minimum the following:
(1) The normal location of
the system's radiation port, the general direction of the useful beam, the
location of any windows and doors or other openings, the location of the
operator's booth, and the location of the control panel;
(2) The structural
composition and thickness or lead equivalent of all walls, doors, partitions,
floor, and ceiling of the room concerned;
(3) The dimensions of any
room concerned;
(4) The type of occupancy
of any adjacent areas inclusive of space above and below the room concerned. If
there is an exterior wall, the plans must show distance to the closest area
where it is likely that individuals may be present;
(5) The make and model of
the equipment, the maximum technique factors, and the energy waveform (single
phase, three phase, etc.); and
(6) The type of
examinations or treatments which will be performed with the equipment and the
anticipated workload of the system in mA-minutes per week.
The department may require the
applicant to utilize the services of a health physicist to determine the
shielding requirements prior to the plan review and approval. The approval of
such plans may not preclude the requirement for additional modifications should
a subsequent change of operating conditions create the possibility of an
individual receiving a dose in excess of the limits.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-17.
Rule 44:03:01:11.02 Installation requirements of all radiation equipment.
44:03:01:11.02. Installation requirements of all radiation
equipment. No person may make, sell, lease, transfer, lend, assemble, or
install any radiation machine or any supplies used in connection with such a
machine unless such supplies and equipment meet the requirements of this
chapter when placed in operation.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-17.
Rule 44:03:01:11.03 Notification of installation of radiation equipment.
44:03:01:11.03. Notification of installation of radiation
equipment. Any assembler who installs radiation machines in this state
shall report to the department within 15 days the following information:
(1) The name and address of
the persons who have received these machines;
(2) The manufacturer,
model, and serial number of each radiation machine and major components
transferred;
(3) The date of transfer of
each radiation machine; and
(4) The date of
installation of each radiation machine.
In the case of diagnostic X ray
systems which contain certified components, a copy of the assembler's report,
Form FDA 2579 (6/95), prepared in compliance with requirements of the Federal
Diagnostic X Ray Standard, 21 C.F.R. 1030.30(d), as of January 1, 1998, shall
be submitted to the department within 15 days following completion of the
assembly.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-17.
Rule 44:03:01:11.04 Licensee requirements after installation of radiation equipment.
44:03:01:11.04. Licensee requirements after installation of
radiation equipment. After completion of construction and installation of
radiation equipment, the licensee shall maintain for inspection:
(1) Model and serial number
of all major components and user's manuals for those components;
(2) Records of surveys,
calibrations, maintenance, and modifications performed on the X ray
systems;
(3) The maximum rated
technique factors of each machine with tube rating and cooling curves charts;
and
(4) Scale drawing of the
room in which a stationary radiation machine system is located with such
drawing indicating the use of areas adjacent to the room and an estimation of
the extent of occupancy by an individual in such areas; the results of a survey
for radiation levels present at the operator's position and at pertinent points
outside the room; and the type and thickness of materials, or lead equivalency,
in each protective barrier used in the room.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-1-17, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-17.
Rule 44:03:01:11.05 Installation, maintenance, and operation of radiation equipment.
44:03:01:11.05. Installation,
maintenance, and operation of radiation equipment. Equipment shall be
installed, maintained, and operated according to the manufacturer's
specifications.
Source:
31 SDR 62, effective November 7, 2004.
General
Authority: SDCL 34-1-17, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-17.
Rule 44:03:01:12 Repealed.
44:03:01:12. X ray machine calibration exposure surveys,
posted date, waivers.Repealed.
Source:
SL 1975, ch 16, § 1; repealed, 6 SDR 93, effective July 1, 1980.
Rule 44:03:01:12.01 Radiation producing equipment calibration.
44:03:01:12.01. Radiation
producing equipment calibration. The licensee shall provide that
calibrations are performed on a diagnostic radiographic system if that system
does not meet the minimum performance criteria specified in
§§ 44:03:01:08.02 to 44:03:01:08.07, inclusive, and if there is any change
or replacement of components that could cause a change in the radiation output
of that system.
The calibration may not exceed three
months after any change or replacement of components that could cause a change
in the radiation output. The calibration of the radiation output of the
X ray system shall be performed by or under the direction of a qualified
expert. Calibration of the radiation output of an X ray system shall be
performed with a calibrated dosimetry system. Any X ray machine shall be
calibrated at least every three years unless it meets the standards of this
chapter. Any computed tomography system shall be calibrated or surveyed by a
medical physicist on an annual basis.
Source:
26 SDR 96, effective January 23, 2000; 32 SDR 128, effective January 30, 2006.
General
Authority: SDCL 34-1-17, 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:12.02 Surveys of radiation producing facilities and radiation equipment.
44:03:01:12.02. Surveys of radiation producing facilities and radiation
equipment. Any new radiation producing facility and any existing radiation
producing facility shall have a survey made by a qualified expert or the
department. The survey shall be done after any change in the facility or
equipment that might cause a significant increase in radiation hazard. The
licensee shall obtain a written report of the survey from a qualified expert.
The licensee shall transmit a copy of the report to the department within 30
days of receipt of the report. The survey and report shall indicate all
instances where the installation, in the opinion of the qualified expert or
department, is in violation of the regulations.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:13 Transferred.
44:03:01:13. Transferred to § 44:03:01:08.01(4).
Rule 44:03:01:14 Repealed.
44:03:01:14. X ray equipment to be operated by trained
individuals.Repealed.
Source:
SL 1975, ch 16, § 1; repealed, 6 SDR 93, effective July 1, 1980.
Rule 44:03:01:14.01 Operator requirements for X ray equipment.
44:03:01:14.01. Operator requirements for X ray equipment.
Any person who is certified by the American Registry of Radiological
Technologists, by the American Registry of Clinical Radiological Technologists,
by another state, or who has documented 40 hours of orientation and training in
the operation of radiation producing equipment by a qualified instructor may
operate any radiation producing device. Dental
radiographers shall have a minimum of 16 hours of training.
Source:
26 SDR 96, effective January 23, 2000; 31 SDR 62, effective November 7, 2004.
General Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-15.
Rule 44:03:01:14.02 Operator training requirements for diagnostic radiation equipment.
44:03:01:14.02. Operator training requirements for diagnostic
radiation equipment. A qualified instructor must do all training for
operators of diagnostic radiation equipment. Documentation of the training must
include the dates, instructor, and subjects covered. Continuing education
credits would qualify as part of the training. The following are areas in which
an individual must have documented training for the operation of X ray
equipment:
(1) Fundamentals of
radiation safety must cover characteristics of radiation, units of radiation
measurement, hazards of exposure to radiation, levels of radiation from
sources, and methods of controlling radiation dose;
(2) Familiarization with
equipment must cover identification of controls, function of each control, how
each control affects technique chart, and utilization of technique chart;
(3) Film processing must
cover film speed as related to patient exposure, film processing with automatic
processors, film processing manually, and factors affecting film processing
quality;
(4) Anatomy and positioning
relative to scope of practice, including patient preparation, and correct
method for performing procedures; and
(5) The requirement of
federal and state regulations pertinent to the services offered.
Source:
26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-14.
Rule 44:03:01:14.03 Operator continuing education requirements.
44:03:01:14.03. Operator continuing education requirements.
Any operator of a radiation producing device shall have five hours of
documented continuing education over a three-year period containing information
on radiation safety, equipment operation, film processing, emergency
procedures, anatomy, positioning of film and body parts, orientation or
training in new developed procedures, infection control, or rules pertinent to
the services offered. Excluded from the five hours of continuing education are
any licensed practitioner of the healing arts and any employee of a dental
facility.
Source:
26 SDR 96, effective January 23, 2000; 31 SDR 62, effective November 7, 2004.
General Authority: SDCL 34-21-4.1, 34-21-15.
Law
Implemented: SDCL 34-21-4.1, 34-21-14.
Rule 44:03:01:15 Disposal of radioactive materials.
44:03:01:15. Disposal of radioactive materials. No
licensee may dispose of any medical radioactive material without prior written
approval of the department unless under license of the Nuclear Regulatory
Commission. Data regarding the potential radium or other radiation hazard are
required.
Source:
SL 1975, ch 16, §1; 6 SDR 93, effective July 1, 1980; 26 SDR 96, effective
January 23, 2000.
General
Authority: SDCL 34-21-15.
Law
Implemented: SDCL 34-21-31.
Rule 44:03:01:16 Loss or theft of radioactive material.
44:03:01:16. Loss
or theft of radioactive material. Each licensee shall
report by telephone fax, or electronic mail and
followed up with a written report to the department the loss or theft of any
source of radiation immediately after the occurrence becomes known.
Source:
SL 1975, ch 16, § 1; 6 SDR 93, effective July 1, 1980; 26 SDR 96,
effective January 23, 2000.
General
Authority: SDCL 34-1-17, 34-21-15.
Law
Implemented: SDCL 34-21-11.
Rule 44:03:01:17 Amendment, suspension, or revocation of license.
44:03:01:17. Amendment, suspension, or revocation of license.
Any amendment to a license shall be made by issuing a new license. Any
suspension or revocation proceeding, in accordance with SDCL chapter 1-26, can be initiated only after exposure surveys are conducted by the department and a serious hazard to public health and safety is determined.
Source:
6 SDR 93, effective July 1, 1980; 26 SDR 96, effective January 23, 2000.
General
Authority: SDCL 34-1-17, 34-21-15.
Law
Implemented: SDCL 34-21-19.
Back to
44 |
44:03 